Water Modifications

Endoscope Water Modification

For DV2 System ONLY

The water flow into the subgingival environment is critical to Perioscope vision subgingivally.  The original DV2 system of water delivery can be unreliable for varying reasons.  Perioscopy Incorporated recommends a “back up” or “by-pass” system that is a simple air pressure driven water bottle modification. This water bottle modification can be made by a competent dental repair technician.

HOWEVER we recommend:

A “plug and play” version can be purchased from:

Danville Materials LLC

3420 Fostoria Way, A200

San Ramon, CA 94583

(925) 973-0710 x205

Attention: Stewart Denham

 

TO CLARIFY: THE DENTALVIEW DESIGN FOR WATER DELIVERY RELIED UPON DENTAL UNIT WATER PRESSURE WHICH IS VARIABLE AND TYPICALLY TOO LOW FOR ADEQUATE VISIBILITY. THIS SYSTEM MODIFICATION RELIES UPON AIR PRESSURE TO PROVIDE WATER FLOW. ADDITIONALLY THE VALVE AND FILTER INSIDE THE MCU HAVE BEEN FOUND TO BE PROBLEMATIC.

WE RECOMMEND 60-65 PSI.

YOUR DENTALVIEW PEDAL WILL ONLY BE USED TO CHANGE THE ILLUMINATION.

THE WATER MODIFICATION HAS ITS OWN PEDAL.

IT ALSO HAS ITS OWN AIR LINE, SIMPLY REQUIRES AN AIR QUICK CONNECT, AND THEN JUST ADD WATER!